Complex of Cations on Primesep C is Effected by Organic Modifier

https://sielc.com/Application-Complex-of-Cations-on-Primesep-C-Is-Effected-By-Organic-Modifier
⬇ PDF Download

Chromatogram

Description

Primesep C separates a mixture of inorganic cations, potassium, sodium, and lithium, by a combination of cation exchange and complex formation. The peak order is reversed on the Primesep C column compared to the order normally obtained on columns that rely on cation exchange alone. In addition, the elution order can be rearranged by simply changing the organic modifier in the mobile phase from acetonitrile to methanol. The mobile phase mixture of water, acetonitrile (MeCN, ACN) or methanol (MeOH) and ammonium acetate is compatible with UV and evaporative light scattering detection (ELSD).

Method Parameters

Mobile PhaseMeOH, MeCN
BufferAmAc pH 5.0
Flow Rate1.0 ml/min
DetectionELSD 50C
Class of CompoundsIons,  Hydrophilic, Ionizable, Vitamin, Supplements
Analyzing CompoundsSodium, Potassium,  Lithium, Chloride

HPLC Column Used

Primesep C, 4.6×50 mm, 5 µm, 100A