Complex Mixture of Acids, Bases, Amino Acids, and Neutral Compounds

https://sielc.com/Application-Complex-Mixture-of-Acids-Bases-Amino-Acids-and-Neutral-Compounds
⬇ PDF Download

Chromatogram

Description

Primesep 100 separates a mixture of amino acids (tyrosine, phenylalanine), organic acids (benzoic acid, mandelic acid), amines (benzylamine, pyridine), and neutrals (benzonitrile, toluene) in one HPLC run by combining reversed-phase, cation-exchange, and polar interactions. The method is tunable and peak order can be changed significantly by adjusting acetonitrile and trifluoroacetic acid concentrations. The separation method uses a mobile phase mixture of water, acetonitrile (MeCN, ACN) and trifluoracetic acid (TFA) and compatible with UV, mass spec (LC/MS) and evaporative light scattering (ELSD) detection.

Method Parameters

Mobile PhaseMeCN/H2O – 30/70%
BufferTFA – 0.2
Flow Rate1.0 ml/min
DetectionUV, 210 nm
Class of CompoundsDrug, Acid, Hydrophilic, Ionizable, Vitamin, Supplements, Amino acid
Analyzing CompoundsTyrosine, phenylalanine, Benzoic acid, mandelic acid, Benzylamine, Pyridine, Benzonitrile, Toluene

HPLC Column Used

Primesep 100, 4.6×250 mm, 5 µm, 100A