Separation of Guanosine Diphosphate Mannose and Guanosine Diphosphate Fucose in Mixed-Mode Chromatography

HPLC-Applikation • Datum: 2013/12/05
Separation of Guanosine Diphosphate Mannose and Guanosine Diphosphate Fucose in Mixed-Mode Chromatography Chromatogramm

Separation type: Liquid Chromatography Mixed-mode

High Performance Liquid Chromatography (HPLC) Method for Analysis of Guanosine Diphosphate Mannose and Guanosine Diphosphate Fucose. Two nucleotides containing different sugar fragments were separated on Primesep SB mixed-mode HPLC column using LC/MS compatible conditions. Method can be used for analysis and isolation of guanosine diphosphate mannose and guanosine diphosphate fucose. Mechanism of retention is weak reversed-phase and strong anion-exchange. retention is controlled by buffer concentration and buffer pH. Both compounds exhibit strong interaction providing good retention and separation.

Methodenparameter

ColumnPrimesep SB, 4.6×50 mm, 5 µm, 100 Å
Mobile PhaseMeCN/H2O – 20/80%
BufferAmFm pH 2.9- 50 mM
Flow Rate1.0 mL/min
DetectionUV, 250 nm