HPLC – MS Method for Analysis of Nimesulide on Newcrom R1 Column

HPLC-Applikation • Datum: 2023/10/04
HPLC – MS Method for Analysis of  Nimesulide on Newcrom R1 Column Chromatogramm

Nimesulide is a non-steroidal anti-inflammatory drug (NSAID) that possesses analgesic and antipyretic properties. Below is some information about Nimesulide:

Nimesulide can be retained, separated and analyzed on a Newcrom R1 reverse phase stationary phase column using an isocratic analytical method with a simple mobile phase of water, Acetonitrile (MeCN), and a ammonium format as a buffer. This analysis method can be detected using UV at 300 nm, an Evaporative Light Scattering Detector (ELSD), or any other evaporative detection method (CAD, ESI-MS)

Methodenparameter

ColumnNewcrom R1, 2.1 x 100 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN – 50%,
BufferAmmonium Formate pH 3.0-20 mM
Flow Rate0.2 mL/min
DetectionUV 300 nm, SIM307.2-