Separation type:Liquid Chromatography Mixed-mode SIELC Technologies
trans-Zeatin is a type of cytokinin , a class of plant hormones that promote cell division and growth, delay aging, and play a crucial role in plant development. It is one of the most active and common forms of zeatin found in plants.
It is often involved in various physiological processes, including seed germination, leaf formation, and root and shoot differentiation.
trans-Zeatin can be retained, separated and analyzed using a Primesep 100 column. The analysis employs an isocratic method with a simple mobile phase comprising water, acetonitrile (MeCN) and Sulfuric acid as a buffer. This method allows for detection using Vis 200 nm.
You can find detailed UV spectra of trans-Zeatin and information about its various lambda maxima by visiting the following link .
| Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 Å, dual ended |
|---|---|
| Mobile Phase | MeCN/H2O- 20/80% |
| Buffer | H2SO4 – 0.1% |
| Flow Rate | 1.0 mL/min |
| Detection | UV 202 nm |
| Limit of Detection | 50 ppb |
| Sample | 1 mg/ml |