HPLC Method for Analysis of Pyridine on Primesep 100 Column

HPLC-Applikation • Datum: 2024/08/12
HPLC Method for Analysis of Pyridine on Primesep 100 Column Chromatogramm

Separation type: Liquid Chromatography Mixed-mode SIELC Technologies

Pyridine is an aromatic heterocyclic organic compound with the chemical formula C5H5N. It is structurally related to benzene, with one methine group (=CH−) replaced by a nitrogen atom.

Pyridine can be retained, separated and analyzed using a Primesep 100 mixed-mode stationary phase column. The analysis employs a gradient method with a simple mobile phase comprising water, acetonitrile (MeCN), and sulfuric acid as a buffer. This method allows for detection using UV 250 nm.

You can find detailed UV spectra of Pyridine and information about its various lambda maxima by visiting the following link .

Methodenparameter

ColumnPrimesep 100, 4.6 x 150 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN – 70%
BufferH2SO4 – 0.05%
Flow Rate1.0 mL/min
DetectionUV 250 nm
Limit of Detection5 ppb