HPLC Method for Analysis of Crystal Violet on Primesep 100 Column

HPLC-Applikation • Datum: 2023/08/17
HPLC Method for Analysis of Crystal Violet on Primesep 100 Column Chromatogramm

Separation type: Liquid Chromatography Mixed-mode

Crystal Violet, also known as Methyl Violet 10B, is a synthetic dye belonging to the triarylmethane dye family. It has the molecular formula C25H30ClN3. This dye is primarily known for its use in microbiology as a stain, but it has other applications as well.

Crystal Violet can be retained, and analyzed on a Primesep 100 mixed-mode stationary phase column using an isocratic analytical method with a simple mobile phase of water, Acetonitrile (MeCN), and a sulfuric acid (H2SO4) buffer. This analysis method can be detected in the UV-Vis regime at 540, 590, and 200 nm.

Methodenparameter

ColumnPrimesep 100, 4.6 x 150 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN/H2O – 80/20%
BufferH3PO4 – 0.2%
Flow Rate1.0 mL/min
DetectionVis, 590 nm
Limit of Detection10 ppb
Injection Volume5 µl