HPLC Method For Analysis Of Alprazolam on Newcrom R1 Column

HPLC-Applikation • Datum: 2018/05/22
HPLC Method For Analysis Of Alprazolam on Newcrom R1 Column Chromatogramm

High Performance Liquid Chromatography (HPLC) Method for Analysis of Alprazolam

Alprazolam, also known as Xanax, is a perscription medication with the C17H13ClN4molecular formula. It is typically used to treat anxiety and panic disorders. It works through enhancing the effects of GABA, which is a calming neurotransmitter in the brain.

Alprazolam can be retained and analyzed using the Newcrom R1 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a phosphoric acid buffer. Detection is performed using UV.

Methodenparameter

ColumnNewcrom R1, 3.2 x 100 mm, 5 µm, 100 Å, dual ended
Mobile PhaseMeCN/H2O – 30/70%
BufferH3PO4 – 0.2%
Flow Rate0.5 mL/min
DetectionUV, 250 nm